  Macurco MA-ND1 Gas Detectors Cilibration Kit

Macurco MA-ND1 Gas Detectors Cilibration Kit


 

Our Price : $884.99


Product Code: MA-ND1

Qty:

Description
 
The ND-1 is a microcomputer controlled electronic sensor and controller for Nitrogen Dioxide (NO2). The unit has a NO2 sensor (electrochemical), digital display, status lights, buzzer and relays. Please see the ND-1 data sheet for more detailed information.

Location:
The unit, on average, can cover about 5000 sq. ft. (465 sq. meters). The coverage depends on air movement in the room or facility. Extra detectors may be needed near any areas where people work, or the air is stagnant. Normally, the unit mounts 5 feet (1.5 meters) above the floor in a central area where air movement is generally good.

Installation:
The ND-1 is housed in an electrical box that has mounting holes on the back. Conduit can be connected to the bottom three knock-out holes. Do NOT remove the four black vents in the case. They are necessary for adequate air flow into the unit. Remove the two small screws toward the bottom of the inside cover. The inside cover then swings up for access to internal circuitry.

Once the cover is open, you should notice a seven-screw terminal block. The first two screws, at the far left, are for power and are marked “POWER 12-24 V ac/dc”. There is no polarity. The three middle terminals are for the Fan SPDT relay. They are marked from left to right N.O. (normally-open), COM (common), and N.C. (normally-closed). The Fan relay contacts can handle up to 5 amps. The two terminals at the far right are for the Alarm relay. They are both marked N.O. (normally-open). The Alarm relay contacts can handle up to 0.5 amps.

The relays have dry contacts. See the ND-1 data sheet for detailed schematics on wiring and other technical information.

Operation:
When power is first applied to the detector, it will go through a warm-up period of 30 seconds. The digital display will count down from 30 to 0 during the warm-up period. The Fan and Alarm relays will also actuate during the warm-up period.

The ND-1 is now operational. Set the Fan and Alarm PPM levels as needed for the job requirements. The current OSHA and ACGIH values for NO2 are TWA of 3 ppm (parts per million) and STEL of 5 ppm. The suggested settings to meet those limits are a Fan setting of 3 ppm and an Alarm setting of 5 ppm.

The Fan turn-on delay can be set from 0 to 5 minutes by setting the 3-segment “FAN DELAY” switch (inside the unit just below the big blue buzzer). When the turn-on delay has been actuated by the NO2 level exceeding the threshold settings, the amber light will flash. When the turn-on delay is over the amber light will come on solid and the Fan relay will actuate.

The Alarm turn-on delay (flashing red) is fixed at 5 minutes. A solid red light indicates Alarm relay activation, and the buzzer will sound. Both Fan and Alarm relays have a minimum on-time of 30 seconds. This is to prevent short cycling devices connected to the relays.

Testing:
Testing the unit can be done by pressing the “TEST” switch located near the buzzer on the printed circuit board. This will initiate the test cycle. The display will read 88.8, both the Fan and Alarm relays will actuate, both Fan and Alarm status lights will be on solid, and the buzzer will sound. The test cycle will last approximately 30 seconds.

Reset:
The reset feature is activated by pressing the switch labeled “RESET”. The switch is mounted on the printed circuit board near the terminal block. Pressing the switch will initiate the warm-up cycle as mentioned above in “Operation”.

Stabilization Time:
Although it will be operational within a few minutes of being first powered, the gas sensor in the ND-1 requires one hour to reach its final operating point. The ND-1 is intended to be continuously powered. If you are certain that the air is clean and the ND-1 indicates a small reading, the zero potentiometer (just above the sensor labeled “ZERO”) may be adjusted. Turn the screw about a quarter of a turn CCW at a time, while waiting 30 seconds between turns. This is a 25-turn potentiometer.

Alarms Due to Other Gases:
Although the ND-1 is specifically calibrated with NO2, the electro-chemical sensor will detect a variety of gases such as Hydrogen Sulfide, Sulfur Dioxide, Chlorine, and Hydrogen Chloride. These gases may cause a reading and relay actuations.

Replacing The Sensor And Recalibrating The Nd-1:
Macurco provides a replacement sensor and a Field Calibration Kit for the ND-1. We recommend that the sensor be replaced, and the unit be re-calibrated at least once every two years. The Field Calibration Kit can be used to calibrate several units. The number of units that can be calibrated is determined by the gas quantity in the gas cylinder. This will vary due to calibration kit useage and gas flow rate. However, the gas mixture has a useful life of no more than one year.

Servicing Of Unit:
The ND-1 does not require regular maintenance. The unit uses a self-purging electro-chemical sensor that has a two year life expectancy. This sensor acts somewhat like a battery in that its output will be reduced with time. The manufacturer of the sensor (City Technology Limited) indicates the sensor can be expected to have a reduced output of up to 2% per month, but not to normally exceed 5% per year. All maintenance and repair of products manufactured by Macurco, Inc. are to be performed at the Macurco manufacturing facility. Macurco does not sanction any third-party repair facilities.
Features
  • Automatic Ventilation Controller
  • Dual relays: Fan (SPDT) & Alarm (N.O.)
  • Microcomputer Controlled
  • Digital Display: 0 to 9 PPM NO2
  • Highly Linear Electrochemical Sensor
  • Selectable Set Points & Delay Time
  • Diesel Exhaust Detection and Control
  • Internally adjustable set points for fan and alarm relays.
  • Meets the requirements of OSHA and the state of Wisconsin.
  • Housed in Nema 1, UL listed, metal knockout case.
  • Adjustable switch for fan delay from 0.5 to 5 minutes.

Share your knowledge of this product with other customers... Be the first to write a review

Browse for more products in the same category as this item:

Shop Brands > 3M Macurco Carbon Monoxide & Fixed Gas Detectors & Monitors
 
 

Enter your email for a chance to WIN $100 Amex Gift Card.

Please Enter Your Email Address

*One entry per household please